Method Development and Validation for Simultaneous Estimation of Esomeprazole and itopride Hydrochloride in Combined Pharmaceutical Dosage Form and bulk by RP- HPLC

 

Sowjanya Vadrevu

Block 1, Building 18, Al Farwaniya, Kuwait.

*Corresponding Author E-mail: vvlsps999@gmail.com

 

ABSTRACT:

The developed RP- HPLC method allows rapid and precise determinations of simultaneous estimation of esmoprazole and itopride hydrochloride in combined pharmaceutical dosage forms. The objective of the proposed method is to develop simple and accurate methods for the determination of simultaneous estimation of esmoprazole and itopride hydrochloride in combined pharmacetcal dosage forms and it’s stability indicative studies. A series of mobile phases were tried, among the various mobile phases, Separation of the drug was achieved on ACE C18 (150×4.6) mm, 5μ using mobile phases, (A) 6.5 pH di hydrogen ortho phosphate buffer, (B) Acetonitrile and Water in the ratio of 35:65v/v. The flow rate was 1.2 ml/min and the detection wavelength was 272 nm.The Linearity and correlation coefficient of esomeprazole and itopride hydrochloride at 10-50µg/ml and 10-50µg/ml was found to be 0.999 and 0.999 respectively. The LOD was found to be 0.57 mg/ml and 0.56mg/ml and LOQ was found to be 1.74mg/ml and 1.69 mg/ml for Esomeprazole and itopride hydrochloride respectively which represents that sensitivity of the method is high. The method was known to be accuratewith the assay method. The % assay was found to be 99.58 % and 99.98%. The developed method was showed to a good accuracy and precision. The % RSD is for Esomeprazole and itopride hydrochloride is 0.55 and 0.75. This method shows that good reproducibility of the results. Furthermore this method was simple, sensitive, and accurate. This method can be employed for routine quality control analysis of tablets of esomeprazole and itopride hydrochloride in combined pharmacetcal dosage forms in various educational institutions.

 

KEYWORDS: Esomeprazole, itopride hydrochloride, Optimization, RP-HPLC.

 

 

 

 

INTRODUCTION:

Esomeprazole (esomeprazole) is a proton pump inhibitor that decreases the amount of acid produced in the stomach. esomeprazole is used to treat symptoms of gastroesophageal reflux disease (gerd) and other conditions involving excessive stomach acid such as zollinger-ellison syndrome. Itopride hydrochloride hasanticholinesterase (AchE) activity as well as dopamine D2 receptor antagonisticactivity. It is well established that M3 receptors exist on the smooth muscle layerthroughout the gut and acetylcholine (ACh) released from enteric nerve endingsstimulates the contraction of smooth muscle through M3 receptors. Several methods have been developed using various chromatographic studies and the scope of the present work is to expand and optimization of the chromatographic conditions, to develop RP-HPLC method. A series of mobile phases were tried, among the various mobile Separation of the drug was achieved on ACE C18 (150×4.6) mm, 5μ using mobile phases, (A) 6.5 pH di hydrogen ortho phosphate buffer, (B) Acetonitrile and Water in the ratio of 35:65v/v.,since it gave a good resolution and peak shapes with perfect optimization. These drugs are evaluated for linearity, precision, accuracy, LOD, LOQ, Specificity, %assay, system suitability, robustness and ruggedness.

 

HIGH PERFOMENCE LIQUIID CHROMATOGRAPHY1-4:

HPLC is a type of liquid chromatography that employs a liquid mobile phase and a very divided stationary phase. In order to obtain satisfactory flow rate liquid must be pressurized to a few thousands of pounds per square inch.

 

The rate of drugs between stationary and mobile phase is controlled by diffusion process, if diffusion is minimized, a faster and effective separation can be achieved. The technique of liquid chromatography is called because of its improved performance in compared to classic column chromatography. Advanced in column technology, high- pressure pumping system and sensitive detectors have transformed liquid column chromatography into high speed, efficient, accurate and highly resolved method of separation.

 

HPLC is the method of choice in the field of analytical chemistry, since this method is specific, robust, linear, precise and accurate and the limit of detection is low.

 

Advantages of HPLC:

·       Speed (many analysis can be accomplished in 20min orless)

·       Grater sensitivity (various detectors can beemployed)

·       Improved resolution (wide variety of stationaryphases)

·       Reusable column (expensive columns but can be used for manyanalysis)

·       Ideal for the substances of lowviscosity

 

The various components of a HPLC system are here with described.

 

Fig No -1: Diagram of HPLC System

 

The main objective of the proposed method is to develop simple and accurate methods for the determination of esmoprazole and itopride hydrochloride in combined pharmaceutical dosageforms.

 

MATERIALS AND METHODS:5-8

Instruments:

Waters 2695 HPLC, UV visible double beam spectrophotometer labindia UV-300, pH1500 Eutech electronics limited SE 60 US ultra sonicator, melting point apparatus.

 

Reagents and chemicals used:

Potassium Di hydrogen ortho phosphate , Acetonitrile, Milli –Q water, Methanol (99.9%) HPLC grade obtained from fischer scientific labs Drugs Used: Esmoprazole and itopride hydrochloride are obtained from krish care laboratories.

 

METHOD DEVELOPMENT9-14:

Proper selection of HPLC method development depends upon the nature of the sample, its molecular weight and solubility. For successful method development various Chromatographic parameters such as pH, mobile phase, its composition and proportion, detection wavelength and other factors were exhaustivelystudied.

 

Selection of Chromatographic method:

Preparation of the Esomeprazole and Itopride hydrochloride Standard Solution: Standard Solution Preparation:

Accurately weigh and transfer 20 and 75mg of Esomeprazole and Itopride hydrochloride working standard into a 10mL clean dry volumetric flask add Diluent and sonicate to dissolve it completely and make volume up to the mark with the same solvent. (Stocksolution).

 

Further pipette 1.0ml of Esomeprazole and Itopride hydrochloride of the above stock solution into a 10ml volumetric flask and dilute up to the mark with diluent.

Further pipette 1.5ml of Esomeprazole and Itopride hydrochloride of the above stock solution into a 10ml volumetric flask and dilute up to the mark with diluent Initialization of the instrument

 

The HPLC instrument was switched on. The column was washed with HPLC water for 45 minutes. The column was then saturated with mobile phase for 45 minute. The mobile phase was run to find the peaks. After 20 minutes the standard drug solution was injected inHPLC.

 

Different chromatographic conditions used and their Optimizations

The different HPLC chromatographic conditions were used to find out the optimum chromatographic condition for best elution of drug

 

TRAIL AND ERROR METHOD:

Six number of trials were conducted and the results are Based up on the results the optimized method was determined.

 

OPTIMIZED METHOD FOR HPLC CHROMATOGRAPHIC CONDITIONS:

Chromatogram was shown in Fig.No. 2 and the parameters are shown in Table .1 and results in Table.2

 

Table.1. Optimized chromatographic conditions

Parameters

Method

Stationary phase (column)

Inertsil -BDS C18(250 x 4.6 mm, 5 µ)

Mobile Phase

buffer: acn (35:65)

Flow rate (ml/min)

1.2 ml/min

Run time (minutes)

10 min

Column temperature (°C)

250c

Volume of injection loop (l)

20 µL

Detection wavelength (nm)

272 nm

Drug RT (min)

Eso is2.1 snd itopride hydrochloride is

 

ANALYTICAL METHOD VALIDATION:

The following parameters were considered for validating the developed method as per ICH guidelines.

 

SYSTEM SUITABILITY:

A Standard solution was prepared by using esomeprazole and itopride hydrochloride working standards as per test method and was injected Five times into the HPLC system.

 

The system suitability parameters were evaluated from standard chromatograms by calculating the % RSD from five replicate injections for esomeprazole and itopride hydrochloride, retention times and peak areas.

 

Acceptance criteria:

1.     The % RSD for the retention times of principal peak from 5 replicate injections of each Standard solution should be not more than 2.0%

2.     The % RSD for the peak area responses of principal peak from 5 replicate injections of each standard Solution should be not more than 2.0%.

3.     The number of the oreticalplates (N) for the esomeprazoleandi to pride hydrochloride peaks is NLT3000.

4.     The Tailing factor (T) for the esomeprazole and itopridehydrochloride peaks is not more then 2.0

 

Observation:

The %RSD for retention times and peak areas were found to be within the limit. Refer Table: 3.

 

SPECIFICITY:

esomeprazole and itopride hydrochloride standard and sample solutions were prepared as per the test method are injected into chromatographic system.

 

Acceptance criteria:

Chromatograms of standard and sample should be identical with near Retention time.

 

Observation:

The chromatograms of standard and sample were same identical with same retention time.

 

PRECISION:

Repeatability:

System precision: esomeprazole and itopride hydrochloride Standard solution prepared as per test method and injected five times.

 

Method precision: esomeprazole and itopride hydrochloride Prepared six sample preparations individually using single as per test method and injected each solution.

 

Acceptance criteria:

The % relative standard deviation of individual esomeprazole and itopride hydrochloride , from the six units should be not more than 2.0%.

 

The individual assays of esomeprazole and itopride hydrochloride should be not less than 98% and not more than 102.0%.

 

Observation:

Test results are showing that the test method is precise.as shown in figure 5

 

Intermediate precision (analyst to analyst variability):

To evaluate the intermediate precision (as known as ruggedness) of the method, precision was performed on different day by using different make column of same dimensions. The standard solution was injected for six times and measured the area for all six injections in HPLC. A study was conducted by two analysts as per test method the %RSD for area of six replicate injections was found to be within the specified limits

 

Acceptance criteria:

The individual assays of esomeprazole and itopride hydrochloride should be not less than 98% and not more than 102% and %RSD of assays should be not more then 2.0% by bothanalysts.

 

Observation:

Individual % assays and % RSD of Assay are within limit and passes the intermediate precision, shown in fig.6.

 

ACCURACY (RECOVERY):

Drug Assay was performed in triplicate as per test method with equivalent amount of and esomeprazole and itopride hydrochloride into each volumetric flask for each spike level to get the concentration of Esomeprazole and itopride hydrochloride equivalent to 50%, 100%, and 150% of the labeled amount as per the test method. The average % recovery of esomeprazole and itopride hydrochloride werecalculated.

 

Acceptance criteria:

The mean % recovery of the Esomeprazole and itopride hydrochloride at each spike level should be not less than 98.0% and not more than 102.0% for both the drugs separately.

 

                              Amount found

% Recovery =   ––––––––––––––– × 100

                             Amount added

 

Observation:

The recovery results indicating that the test method has an acceptable level of accuracy.

 

LINEARITY OF TEST METHOD:

A Series of solutions are prepared using Esomeprazole and itopride hydrochloride working standards at concentration levels from 20µg/mL to 80µg/mL of target concentration .Measure the peak area response of solution at Level 1 and Level 6 six times and Level 2 to Level 5 two times.

 

Linearity:

Linearity solutions were prepared such that 0.2mL, 0.3 mL, 0.4mL, 0.5mL, 0.6mL, 0.7mL, 0.8mL from the stock solutions esomeprazole and itopride hydrochloride were taken in to 7 different volumetric flasks and diluted to 10mL with diluents to get 20µg/mL, 30µg/mL, 40 µg/mL, 50µg/mL.

 

Acceptance criteria:

Correlation Coefficient should be not less than 0.9990.

% of y- Intercept should be ±2.0.

 

% of RSD for level 1 and Level 6 should be not more than 2.0%.

 

Observation:

The linear fit of the system was illustrated graphically. The results are shown, Figure; 3 and 4

 

RUGGEDNESS OF TEST METHOD:

a)       System to systemvariability:

System to system variability study was conducted on different HPLC systems, under similar conditions at different times. Esomeprazole and itopride hydrochloride solutions Six samples were prepared and each was analyzed as per test method. Comparison of both the results obtained on two different HPLC systems, shows that the assay test method are rugged for System to system variability.

 

Acceptance criteria:

The % relative standard deviation of Esomeprazole and itopride hydrochloride from the six sample preparations should be not more than 2.0%

 

The % assay of Esomeprazole and itopride hydrochloride should be between 98.0%- 102.0%.

 

Observation:

The % RSD was found within the limit, shown in figure no.7.

 

b)      Column to columnvariability:

Column to column variability study was conducted by using different columns. Esomeprazole and itopride hydrochloride solutions Six samples were prepared and each was analyzed as per test method

 

Acceptance criteria:

The % RSD of Esomeprazole and itopride hydrochloride tablets should be not more than 2.0%.The % assay of Esomeprazole and itopride hydrochloride should be between 98.0% and 102.0% for individual drugs.

 

Observation:

The results obtained by comparing with both two types were within limit.

 

ROBUSTNESS:

a)    Effect of variation of flowrate:

A study was conducted to determine the effect of variation in flow rate. Esomeprazole and itopride hydrochloride Standard solution prepared as per the test method was injected into the HPLC system using flow rates, 1.0 mL/min and 1.2 mL/min. The system suitability parameters were evaluated and found to be within the limits for 1.0 mL/min and 1.2 mL/min flow. Esomeprazole and itopride hydrochloride and was resolved from all other peaks and the retention times were comparable with those obtained for mobile phase having flow rates 1.0mL/min.

 

Acceptance criteria:

The Tailing Factor of Esomeprazole and itopride hydrochloride standards should be not more then 2.0 for Variation inFlow.

 

Observation:

The tailing factor for Esomeprazole and itopride hydrochloride was found to be within the limits, as shown in figure.no.8

 

b)    Effect of variation oftemperature:

A study was conducted to determine the effect of variation in temperature. Esomeprazole and itopride hydrochloride Standard solution prepared as per the test method was injected into the HPLC system at 20şC temperature. The system suitability parameters were evaluated and found to be within the limits for a temperature change of 20şc. Similarly sample solution was chromatographed at 25şC temperature. Esomeprazole and itopride hydrochloride were resolved from all other peaks and the retention times were comparable with those

 

Acceptance criteria:

The Tailing Factor of Esomeprazole and itopride hydrochloride standard and sample solutions should be not more than not more then 2.0 for Variation in temperature.

 

Observation:

The tailing factor for esomeprazole and itopride hydrochloride, x is found to be within the limits,

 

LIMIT OF DETECTION AND QUANTITATION (LOD and LOQ):

From the linearity data calculate the limit of detection and quantitation, using the following formula.

 

LOD:

It is the lowest amount of the analyte in a sample that can be detected but not necessarily be quantitated as an exact concentration or amount

 

LOD= 3.3 σ

              S

LOQ:

It is the lowest amount of an analyte that can be measured quantitatively in a sample with acceptable accuracy and precision. The LOQ is a parameters for tests measuring impurities in the drugproduct.

σ = standard deviation of theresponse

 

S = slope of the calibration curve of the analyte.

 

LOQ = 10 σ

               S

σ = standard deviation of the response

S = slope of the calibration curve of the analyte.

 

DEGRADATION STUDIES:

The International Conference on Harmonization (ICH) guideline entitled stability testing of new drug substances and products requires that stress testing be carried out to elucidate the inherent stability characteristics of the active substance. The aim of this work was to perform the stress degradation studies on the Esomeprazole and Itopride hydrochloride using the proposed method. the results of degradation are shown intable.4

 

Preparation of stock:

Accurately weigh and transfer 15mg and 50mg of Esomeprazole and Itopride hydrochloride working standard into a 10ml clean dry volumetric flask add Diluent and sonicate to dissolve it completely and make volume up to the mark with the same solvent. (Stocksolution)

 

Further pipette 1ml of Esomeprazole and Itopride hydrochloride of the above stock solution into a 10ml volumetric flask and dilute up to the mark with diluent.

 

Hydrolytic degradation under acidic condition:

Pipette 1.5ml of above solution into a 10ml volumetric flask and 3ml of 0.1N HCl was added. Then, the volumetric flask was kept at 60şC for 6 hours and then neutralized with 0.1 N NaOH and make up to 10ml with diluent. Filter the solution with 0.22 microns syringe filters and place in vials.

 

Hydrolytic degradation under alkaline condition:

Pipette 1.5ml of above solution into a 10ml volumetric flask into a 10ml volumetric flask and add 3ml of 0.1N NaOH was added in 10ml of volumetric flask. Then, the volumetric flask was kept at 60şC for 6 hours and then neutralized with 0.1N HCl

and make up to 10ml with diluent. Filter the solution with 0.22 microns syringe filters and place in vials.

 

Thermal induced degradation:

Esomeprazole and Itopride hydrochloride sample was taken in petridish and kept in Hot air oven at 1100C for 24 hours. Then the sample was taken and diluted with diluents and injected into HPLC and analysed.

 

Oxidative degradation:

Pipette 1.5ml above stock solution 2 into a 10ml volumetric flask solution into a 10ml volumetric flask 1 ml of 3% w/v of hydrogen peroxide added in 10ml of volumetric flask and the volume was made up to the mark with diluent. The volumetric flask was then kept at room temperature for 15 min. Filter the solution with 0.45 microns syringe filters and place in vials.

 

RESULTS AND DISCUSSION:

DEVELOPMENT OF METHOD FOR THE ESTIMATION OF ESOMEPRAZOLE AND ITOPRIDE HYDROCHLORIDE7-9:

Several trails were conducted during the development of a method for the simultaneous estimation of Esomeprazole and itopride hydrochloride in bulk form. The best peak was shown in below chromatograms.based on these trials an optimized method was developed and the chromatogram was shownbelow:

 

Fig 2: A typical chromatogram for optimized method

 

Table 2. Optimized method

S. NO

Name of the peak

Retention time(min)

1

Esomeprazole(mg)

2.149

2

itopride hydrochloride

3.135

 

Observation:

This is showing good peak resolution for two samples. For this reason it was taken as a standard method procedure for the subsequent experiment

 

Table no 3: System suitability parameters of Esomeprazole and itopride hydrochloride

Parameters

Escitalopram

Etizolam

Area

187069.4

577513.6

Retention time

2.14

3.13

Theoretical plates

3682

2859

Tailingfactor

0 .93

1.2

 

Fig no: 3 and 4 Linear calibration curve of Esomeprazole and itopride hydrochloride

 

Accuracy:

The mean recoveries were found to be 99.73% for Esomeprazole and 102.0% for itopride hydrochloride. The limit for mean % recovery is 95-105% and as both the values are within the limit, hence it can be said that the proposed method wasaccurate.

 

PRECISION:

Results of system precision for ESOMEPRAZOLE (mg) AND ITOPRIDE HYDROCHLORIDE

 

Method precision (Repeatability):

 

Fig 5: A typic al Chromatograms for precision

 

Fig 6: A typic al Chromatograms for Intermediate precision

 

 

 

Acceptance Criteria:

The % RSD for the area of five standard injections results should not be more than 2%.the results are found to be with in limits.

 

Assay of Esomeprazole and itopride hydrochloride In Dosage Form:

Assay was performed and the results are with in limits and was found to be 99.59 for esomeprazole and 99.98 for itopride hydrochloride.

 

RUGGEDNESS:

a) System to System variability:

 

Fig .7. A typical Chromatograms for system to system variability

 

Robustness:

a) Effect of variation of flow rate for (1mL/min flow)

 

Fig 8: Atypical chromatograms for (1.2 mL/min flow)

 

Acceptance criteria:

The % RSD peak area of all peaks for the four replicates injections should be not more then 2%.

 

Observation:

The % RSD peak area was found to be within limit . So the method was robust.

 

 

 

LIMIT OF DETECTION AND LIMIT OF QUANTITATION (LOD and LOQ): ESOMEPRAZOLE:

From the linearity plot the LOD and LOQ are calculated:

 

LOD = 3.3 σ

               S

 

      3.3×1774.524

= ------------------=0.94

6204

 

LOQ = 10 σ

               S

 

       10×1774.524

=  -----------------=2.86

6204

 

Itopride hydrochloride:

 

LOD = 3.3 σ

                S

 

        3.3×1840.015

= -------------------–=1.19

5097

 

LOQ = 10 σ

               S

 

      10×1840.015

= ---------------––––=3.60

5097

 

The LOD was found to be 0.94 mg/ml and 1.19mg/ml and LOQ was found to be 2.86 mg/ml and 3.60mg/ml for Escitalopram and Etizolam respectively which represents that sensitivity of the method is high.

 

Degradation:

RESULTS:

Table: 4 Table showing results of amount of drug degradation:

 

Esomeprazole

Itopride hydrochloride

Area

% Degraded

Area

% Degraded

Standard

92911

 

394495

 

Acid

87708

5.6

380688

3.5

Base

90867

2.2

362146

8.2

Peroxide

82784

10.9

376743

4.5

Thermal

82970

10.7

357412

9.4

 

Table 5. Summary of validation parameters by HPLC method

 

PARAMETERS

RESULTS

ESOMEPRAZOLE

Itopride hydrochloride

Method precision

%RSD = 0.641801

%RSD = 0.154822

Intermediate precision

%RSD = 0.951074

%RSD =0.318806

Linearity

10-50 µg/ml

correlation coefficient =0.999

10-50 µg/ml

correlation coefficient =0.999

Accuracy

Recovery values =99.71

100% -102%

Recovery values =100.28

97 %-101%

LOD

0.57

0.56

LOQ

1.74

1.69

robustness

0.55

0.75

 

CONCLUSION:

Hence, the chromatographic method developed for Esomeprazole and itopride hydrochloride is said to be rapid, simple, specific, sensitive, precise, accurate and reliable that can be effectively applied for routine analysis in research institutions, quality control department in industries, approved testing laboratories, bio-pharmaceutics and bio-equivalence studies and in clinical pharma cokineticstudies.

 

REFERENCES:

1.      SS Kadam, KR Mahadik, KG Bothra, Principles of Medicinal Chemistry vol. 1, Eighteenth Edition, Nirali Prakashan, Pune, 2007,259-264.

2.      RS Satoskar, SD Bhandarkar, NN Rege; Pharmacology and pharmacotherapeutics, Ninth edition, Popular Prakashan, Mumbai, 1998,241-257

3.      KD Tripathi, Essential of Medical Pharmacology, 5th edition, Jaypee Brothers, Medical Publishers, India, 2004,352-361

4.      HP Rang, NM Dale, JM Ritter, PK Moore, Pharmacology, Fifth edition, Churchil Livingstone, Edinbrugh, 2003,258-263

5.      FSK Barar, Essential of Pharmacotherapeutics, Third Edition, Chand (S.) and Co Ltd, India, 2005, 152-165

6.      AH Beckett, JB Stenlake, Practical Pharmaceutical Chemistry, Fourth Edition, CBS Publishers and Distributors, New Delhi, India, 2004,296-299

7.      G Stephen, G Schulman, BS Vogt, JW Munson, Pharmaceutical Analysis ModernMethods, Part – II, International Medical Book Distributors, Mumbai, India, 2001,101-109

8.      GH Jeffery, J Bassett, J Mendham, RC Denney, Vogel’s Textbook of QuantitativeChemical Analysis, Fifth Edition, Longman Singapore Publishers Pt. Ltd., Singapore,1989,212-235

9.      DA Skoog, FJ Holler, SR Crouch, Principles of Instrumental Analysis, FifthEdition, Southern College Publication, Japan, 2006, 725-728

10.   GR Chatwal, SK Anand, Instrumental method of Chemical Analysis, Fifth Edition, Himalaya Publication House, Mumbai, 2002,2.624-2.631.

11.   JE De Muth, Basic Statistics and Pharmaceutical Statistical Applications, Second Edition, Chapmanand Hall/CRC., New York, 2006, 169–191.

12.   SB Chepurawr, AA Ahirkhedkar, SB Bari, SJ Surana, Indian Drugs, 2006, 43(10), 803-806.

13.   S Gopinath, S Muralidharan, S Rajan and S P Danaphal, Der Pharmacia Lettre, 2009, 1(1), 135-142.

14.   JSK Nagarajan, S Muralidharan, Der Pharmacia Lettre, 2009, 1(1),162-168.

 

 

 

Received on 15.02.2020       Modified on 08.03.2020

Accepted on 28.03.2020      ©Asian Pharma Press All Right Reserved

Asian J. Pharm. Ana. 2020; 10(2):91-98.

DOI: 10.5958/2231-5675.2020.00016.2